Omnigrad S TAF12D
High temperature TC thermometer

Omnigrad S TAF12D
High temperature TC thermometer ©Endress+Hauser

Applicable for ceramic baking ovens, brickworks, porcelain production and glass industries

Measuring points with extremely high process temperatures require these special thermometers with ceramic thermowells and thermocouples made from special metals, such as platinum and rhodium. The thermowell and the internal sheath made from ceramic materials protects from mechanical and chemical damages in the process and increases the life span of the sensors. This leads to: Cost savings for maintenance of the measuring point, quality improvements of the products and increased plant safety.

  • Voordelen

    • Long lifetime by usage of innovative thermowell materials with increased wear and chemical resistance

    • Long term stable measurement due to sensor protection with non-porous materials

    • Flexible product selection by modular design

    • Optimized life cycle costs...

  • Toepassingsgebied

    • The TAF12D are assemblies with double ceramic protection tubes, designed specifically for applications such as ceramic baking ovens, brickworks, porcelain production and glass industries. They contain a single or double TC insert in ceramic insulators.

    • ...

Kenmerken en specificaties

Terug

Details

  • Meetprincipe

    Thermocouple

  • Characteristic / Application

    metric style

    modular temperature assembly

    high temperature probe (furnace)

    incl. double ceramic thermowell

  • Terminal head

    TA20A - aluminium - cover with screws

    TA20D - aluminium - hinged cover high

    TA30A - E+H, aluminium - cover low - w/o. display

    TA30D - E+H, aluminium - cover high

  • Thermowell / protection tube

    ceramic protection tube and inner tube

  • Insert / probe

    ceramic capilarry, rigid

  • Outer diameter protection tube / Insert

    14,0 mm (0,55")

    15,0 mm (0,59")

  • Max. immersion length on request

    up to 3.500,00 mm (137,80'')

  • Material protection tube/ thermowell

    Ceramic (C610/C799)

  • Process connection

    compression fitting:

    G3/4"

    G1"

    G11/2"

    adjustable Flange

  • Tip shape

    straight

  • Operating temperature range

    Type S:

    0 C ...1.600 C

    (32 F ...2.912 F)

    Type R:

    0 C ...1.600 C

    (32 F ...2.912 F)

    Type B:

    600 C ...1.700 C

    (1.112 F ...3.092 F)

  • Max. process pressure (static)

    at 20 °C: 1 bar (15 psi)

  • Accuracy

    class 2 acc. to IEC 60584

  • Integration head transmitter

    yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

  • Certification

    SIL (transmitter only)

Documenten / handleidingen / software

Accessoires / Spare parts

Nr
Order code
Specification
  • C
    • Nr C Order code 60000516

      Stop Flange adjustable,DIN EN 50446,23mm

      Maximum temperature 400 oC. Material cast iron. Counter flangeand gasket not provided. No gas-tight connection. Dimension: d2=23 mm, a=90 mm, c=70 mm. Clampable diameter from 21mm to 22mm.

    • Order code 60020836

      Knelwartel geschr. G3/4 M d.22,5 316Ti

      Maximum temperatuur 350oC Maximum druk 1 bar. Material roestvast staal Afmetingen: A=G 3/4", d=22mm, Wr=41

    • Order code 60021425

      GCP assemblage, gasdicht, G1 1 / 2 ", 22

      Max.temperatuur 350 °C Max. druk 1 bar. Materiaal 316Ti (1.4571). Voor buis diameters 21,3/22mm. Afmetingen: A=G 1 1/2", d=22,5 mm,Wr=55.

    • Order code 60021757

      Geschroefde knelfitting G1" pijp 22mm

      Maximum temperatuur 350oC Maximum pdruk 1 bar. Materiaal RVS Afmetingen: A=G1", d=22mm, Wr=41

    • Order code 71182602

      Spare parts kit sealing TA30A/TA30D

      5x O-ring 3.69x1.78 5x 0-ring 74.7x2.75

    • Order code 71044061

      Fixing kit cpl.TMT82/85/84 (europees)

      2x schroef, 2x veer, 2x springveer, 1x afdichting CDI-connector

  • B
    • Nr B Order code TMT181-

      iTEMP TMT181, koptransmitter PCP

      Temperatuur transmitter, PC- programmeerbaar. Toepassing: RTD, TC, TC, Ohm, mV. 2-draads 4-20mA, galvanische isolatie. Foutreactie: NAMUR NE 43. Montage: kopvorm B, DIN EN50446. UL vermeld.

    • Nr B Order code TMT182-

      iTEMP TMT182, koptransmitter HART

      Temperatuurtransmitter, Protocol HART. Toepassing: RTD, TC, Ohm, mV. 2-draads 4-20mA, galvanische isolatie. Foutreactie: NAMUR NE 43. Montage: kop vorm B, DIN EN50446. UL vermeld. Temperatuurtransmitter, Protocol HART. Toepassing: RTD, TC, Ohm, mV. 2-draads 4-20mA, galvanische isolatie. Foutreactie: NAMUR NE 43. Montage: kop vorm B, DIN EN50446. UL vermeld.

    • Nr B Order code TMT82-

      iTEMP TMT82, Transmitter HART

      2-draads temperatuurtransmitter. Dubbele universele sensoringang. Galvanische scheiding 2 kV (input/ output). Toepassing: RTD, TC, Ohm en mV. :: Zeer betrouwbaar door drift detectie en sensor back-up. :: Optioneel ook voor veiligheidskritieke applicaties door middel van SIL 2/3 acc. naar IEC61508

    • Nr B Order code TMT84-

      iTEMP TMT84, Transmitter PA

      PROFIBUS PA temperatuur transmitter voor kop- of veldmontage met dubbele sensoringang en advanced diagnostics voor sensorbewaking. Profibus PA Profile 3.02. Galvanische isolatie 2kV. Stroomverbruik: 11 mA. Montage: aansluitkop vorm B, DIN EN50446. Standaard fabrieksinstelling: Kanaal 1: Pt100, 3-draads, oC. kanaal 2: inactief. Meerdere configuraties selecteerbaar.

    • Nr B Order code TMT85-

      iTEMP TMT85, Transmitter FF

      FOUNDATION Fieldbus temperatuur transmitter voor kop- of veldmontage met dubbele sensoringang en advanced diagnostics voor sensorbewaking. Galvanische isolatie 2 kV Meetsignaal: RTD, TC, Ohm, mV. Stroomverbruik: 11 mA.

    • Order code FXA195-

      Commubox FXA195 (USB)

      USB HART interface ten behoeve van universele configuratie tools (b.v. FieldCare) De stroomuitgang (4..20mA) is intrinsiek veilig uitgevoerd (SIL2, IEC61508) De interface wordt via de bijgeleverde kabel aan een PC of laptop aangesloten. De interface wordt gevoed via USB

    • Order code TXU10-

      Configuratiekit TXU10

      Set-up software incl. USB- interfacekabel. Dit betreft een product uit het E-direct portfolio waar staffelprijzen aan gekoppeld zijn. Kijk voor meer informatie op www.e-direct.endress.com

    • Order code READWIN-

      ReadWin2000 configuratie/uitleessoftware