Chlorine dioxide CCS240

Dit product is niet langer leverbaar.

Zie de nieuwe generatie in het deel voorganger/opvolger of neem contact op met uw locale Endress+Hauser-vestiging.

Kenmerken en specificaties

Terug

Details

  • Meetprincipe

    Chlorine dioxide

  • Application

    Waste water, pool

  • Characteristic

    Amperometric measurement of dissolved Chlordioxide

  • Measurement range

    0,05 - 20ppm Chlordioxide;

  • Measuring principle

    - closed (membrane covered) measuring cell.
    - Reduction of Chlordioxid (ClO2) to Chlorid at the cathode at appr. 120mV.

  • Design

    Closed amperometric 2-electrode measuring cell with PTFE membrane

  • Material

    Sensor shaft : PVC
    Membrane : PTFE
    Membrane cap : PBT (GF30); PVDF

  • Dimension

    Diameter : appr. 25mm
    Length : 160mm

  • Process temperature

    2°C ... 45°C

  • Process pressure

    0bar - 1bar

  • Temperature sensor

    With NTC-temperature sensor

  • Connection

    Fixed cable

Documenten / handleidingen / software

Accessoires / Spare parts

Nr
Order code
Specification
  • 1
  • 2
  • A
    • Nr A Order code 50005255

      CCY14-WP

      Geschikt voor CCS140/141 of CCS240/241. Set van 2 st. reserve membranen.

  • A
    • Nr A Order code 71076922

      Maintenance kit, CCS24x

      Inhoud: Elektrolyt 50ml/1.69 fl.oz.; 2 membraan caps; polish folie (2 kwaliteiten)

    • Order code 50005181

      AANSLUITDOOS VBC

      Voor kabel verlenging van chloor / dioxide sensor CCS14x / 24x, beschermingsgraad IP65.

    • Order code 50085333

      MEETKABEL CYK71

      Voor pH-/Redox of 2-elektroden geleidbaarheidssensoren met geintegreerde temperatuuropnemer. Prijs per meter (minimale lengte 5 m).

    • Order code 71257946

      fotometer M+N PF-3, desinf. (chlorine)

      colorimetric measurement of free and total chlorine 0,05 .. 6mg/l Disinfection, drinking and process water hand held photometer, IP67-graded, with cuvettes, DPD, according to ISO 7393 deliverey in plastic case, test kit for free and total chlorine included ::hand held photometer for field cal. and verification of desinfection loops

    • Order code COY8-

      Zero-point gel COY8

      Nulpunt: Verificatie, kalibratie, afstelling, afstelling op het nulpunt van zuurstof en desinfectiesensoren ::Amperometrische en optische sensoren ::12, 25 en 40mm diameter ::Toepassing in het labo, in het veld

  • Opvolger