Omnigrad S TR62
Explosion-proof Pt100 Thermometer

Product picture of resistance thermometer TR62 ©Endress+Hauser

Safe monitoring of process temperatures in challenging applications e. g. in the Oil & Gas industry

The robust thermometer is designed for use in demanding and safety relevant applications e.g. in the Chemical, Oil & Gas, Energy industry. The sensor is designed for use in a separate thermowell. An optional head transmitter with all common communication protocols makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. The modular design with a variety of options can be adapted to the measurement task. Flexible configuration possible.

  • Voordelen

    • Types of protection for use in hazardous locations: Intrinsic safety (Ex i.a.), flameproof (Ex d), non-sparking (Ex nA)

    • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion...

  • Toepassingsgebied

    • Heavy duty applications

    • Oil & Gas processing industry

    • Measuring range: -200 to +600 °C (-328 to 1112 °F)

    • Protection class: IP68

Kenmerken en specificaties

Terug

Details

  • Meetprincipe

    Resistance Temperature Detector

  • Characteristic / Application

    US style metrical

    modular temperature assembly

    for heavy duty applications

    suitable for hazardous areas

    with neck

    to be used with thermowell

  • Terminal head

    TA21H – aluminium – screw cover w. chain

    TA21H – stainless steel – screw cover w. chain

    TA30H – aluminium – w. or w/o. display

    TA30H – aluminium – one or two cable entries NPT1/2", NPT3/4", M20x1.5

  • Thermowell / protection tube

    to be used with thermowell

  • Insert / probe

    mineral insulated (MI), flexible

  • Outer diameter protection tube / Insert

    Insert Diameter

    3,0 mm (0,12'')

    6,0 mm (0,24'')

  • Max. immersion length on request

    up to 30.000,00 mm (1.181,10'')

  • Material protection tube/ thermowell

    Not defined

  • Optional coating

    Not defined

  • Process connection

    Thermometer connection

    male thread:

    NPT1/2"

  • Tip shape

    Not defined

  • Operating temperature range

    PT100 TF StrongSens:

    -50 °C ...500 °C

    (-58 °F ...932 °F)

    PT100 WW:

    -200 °C ...600 °C

    (-328 °F ...1.112 °F)

    PT100 TF:

    -50 °C ...400 °C

    (-58 °F ...752 °F)

  • Max. process pressure (static)

    at 20 °C: 500 bar (7.252 psi)

  • Accuracy

    class A acc. to IEC 60751

    class AA acc. to IEC 60751

  • Response time

    depending on configuration

    t50 = 4 s

    t90 = 8 s

  • Integration head transmitter

    yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

  • Ex - approvals

    ATEX II

    IECEx

    NEPSI

    EAC Ex

    Explosion proof

  • Certification

    Gost Metrology

    SIL (transmitter only)

Documenten / handleidingen / software

Accessoires / Spare parts

Nr
Order code
Specification
  • C
    • Order code 71044061

      Fixing kit cpl.TMT82/85/84 (europees)

      2x schroef, 2x veer, 2x springveer, 1x afdichting CDI-connector

  • A
    • Nr A Order code TPR100-

      Meetinzet TPR100

      Pt100 meetinzet, diam. 3 of 6mm, isolatie MgO, RVS mantel. Geconstrueerd vlgs. DIN EN 60751, DIN 43735. Meetbereik afh. van type meetelement.

    • Nr A Order code TPR300-

      Insert TPR300

      Pt100 meetinzet voor de EEx d serie TR6x, diam. 6mm, isolatie MgO, RVS mantel. Geconstrueerd vlgs. DIN EN 60751, DIN43735. Meetbereik afh. van type meetelement: thin film -50 tot +400°C of wire-wound -200 tot +600°C.

    • Nr A Order code TS111-

      Insert iTHERM TS111

      Vervangbaar, metrische versie volgens DIN43735.

  • D
    • Nr D Order code TA21H-

      Aansluitkop TA21H

      "Product wordt afgebouwd - Beperkte beschikbaarheid". Materiaal: aluminium Omgevingstemperatuur: -40 to 100oC Beschermingsklasse IP66/68

  • B
    • Nr B Order code TMT84-

      iTEMP TMT84, Transmitter PA

      PROFIBUS PA temperatuur transmitter voor kop- of veldmontage met dubbele sensoringang en advanced diagnostics voor sensorbewaking. Profibus PA Profile 3.02. Galvanische isolatie 2kV. Stroomverbruik: 11 mA. Montage: aansluitkop vorm B, DIN EN50446. Standaard fabrieksinstelling: Kanaal 1: Pt100, 3-draads, oC. kanaal 2: inactief. Meerdere configuraties selecteerbaar.

    • Nr B Order code TMT85-

      iTEMP TMT85, Transmitter FF

      FOUNDATION Fieldbus temperatuur transmitter voor kop- of veldmontage met dubbele sensoringang en advanced diagnostics voor sensorbewaking. Galvanische isolatie 2 kV Meetsignaal: RTD, TC, Ohm, mV. Stroomverbruik: 11 mA.

    • Nr B Order code TID10-

      Display voor koptransmitter TID10

      Opsteek display voor koptransmitters. Dot-matrix met weergave gemeten waarde. 12 DIP-switches aan onderzijde. Instellingen: FF: Simulatie, write Lock. PA: Bus adres, write Lock. Geschikt voor: TMT85, TMT84 en TMT82

    • Order code READWIN-

      ReadWin2000 configuratie/uitleessoftware

      ReadWin2000 configuratie software

    • Order code TXU10-

      Configuratiekit TXU10

      Set-up software incl. USB- interfacekabel. Dit betreft een product uit het E-direct portfolio waar staffelprijzen aan gekoppeld zijn. Kijk voor meer informatie op www.e-direct.endress.com

    • Order code FXA195-

      Commubox FXA195 (USB)

      USB HART interface ten behoeve van universele configuratie tools (b.v. FieldCare) De stroomuitgang (4..20mA) is intrinsiek veilig uitgevoerd (SIL2, IEC61508) De interface wordt via de bijgeleverde kabel aan een PC of laptop aangesloten. De interface wordt gevoed via USB

  • D
    • Nr D Order code TA30H-

      Aansluitkop TA30H

      Aansluitkop voor aansluitblok of koptransmitter + optioneel display. Deksel: geschroefd, optioneel met venster. Materiaal: aluminium Coating: polyester poeder-coating. Kleur: huis blauw, deksel grijs. Omgevingstemperatuur.: -50...150 oC. Beschermingsklasse: IP66/68, NEMA 4X.

  • B
    • Nr B Order code TMT180-

      iTemp TMT180, koptransmitter Pt100

      Temperatuur transmitter, PC programmeerbaar. Toepassing: OTO. 2-draads 4-20mA. Foutreactie: NAMUR NE 43. Montage: kopvorm B, DIN EN50446. UL vermeld.

    • Nr B Order code TMT181-

      iTEMP TMT181, koptransmitter PCP

      Temperatuur transmitter, PC- programmeerbaar. Toepassing: RTD, TC, TC, Ohm, mV. 2-draads 4-20mA, galvanische isolatie. Foutreactie: NAMUR NE 43. Montage: kopvorm B, DIN EN50446. UL vermeld.

    • Nr B Order code TMT182-

      iTEMP TMT182, koptransmitter HART

      Temperatuurtransmitter, Protocol HART. Toepassing: RTD, TC, TC, Ohm, mV. 2-draads 4-20mA, SIL2, galvanische isolatie. Foutreactie: NAAM NE 43. Montage: kopvorm B, DIN EN50446. UL vermeld.

    • Nr B Order code TMT82-

      iTEMP TMT82, Transmitter HART

      2-draads temperatuurtransmitter. Dubbele universele sensoringang. Galvanische scheiding 2 kV (input/ output). Toepassing: RTD, TC, Ohm en mV. :: Zeer betrouwbaar door drift detectie en sensor back-up. :: Optioneel ook voor veiligheidskritieke applicaties door middel van SIL 2/3 acc. naar IEC61508